사료의 비타민 D2 및 D3 분석방법에 관한 연구 1 . 고성능 액체 크로마토그라프를 이용한 표준곡선 작성
Studies on the Analytical Method of Vitamin D2 and D3 in Feedstuffs 1 . Calibration curve using high performance liquid chromatograph
장윤환 , 이선행 , 여영수 , 전진석 ( Y . H . Chiang , S . H . Lee , Yeoh , J . S . Cheon )
UCI I410-ECN-0102-2008-520-002137461
* This article is free of use.

This study was carried out as a preliminary step for analyzing the vitamin D₂ and D₃ in feedstuffs using HPLC. 1. Absorbance curve following the wavelength of UV light showed that the 264㎚ gave a maximum absorbance. 2. The mixed solution of 20 ppm vitaminD₂ and D₃ was injected in HPLC with μ Bondapak C_(18) column. The vitamin D₂ was resolved at 14.8 min. and D₃ at 15.4 min. after injection when eluted with 95% methanol + 5 % H₂O. 3. Varying amount of vitamin D₂ was injected into the column of μ Bondapak C_(18), and the detector 240 showed that the least detectable level was 5ng and that 5-100ng range brought a linear calibration curve. 4. When μ Bondapak column and 440 detector were used, 10-100ng range gave a linear standard curve. 5. When μ Porasil column and 440 detector were applied, 10-100ng range showed a linear calibration curve.