바이오 시료의 적정온도 사이클 유지를 위한 채널형 히트싱크에 대한 실험적 연구
Experimental Study on the Channel Type Heat Sink to Maintain Proper Temperature Cycle of Bio-Sample
황정규 ( Jeong-gyu Hwang ) , 박상희 ( Sang-hee Park )
UCI I410-ECN-0102-2023-500-001118366

This study was conducted experimentally to investigate the surface temperature of the heat sink, the air temperature in the flow channel and the sample temperature by changing the channel number of channel type heat sink and the air flow rate when heating and cooling the bio sample. The target temperature of the sample was 15℃ or less as the minimum value and 82℃ or more as the maximum value. In this study, the channel number of the heat sink(N = 1, 2, 4, 5, 10) and the air flow rate(Q=25, 42, 54m3/min) were varied. The bio sample was replaced with water, and the volume of water is 4mL. The size of the heat sink is 80x73x150mm and the material is aluminum. When cooling the sample, the surface temperature, the air temperature and the sample temperature were highly dependent on the number of channels and the flow rate. However, when the sample is heated, the surface temperature, air temperature and sample temperature do not depend on the number of channels and the flow rate. It was found that the conditions for satisfying the minimum temperature of 15℃ or less when cooling the sample were the number of channels N≥5 and the flow rate Q≥42m3/min. When heating the sample, the conditions to satisfy the maximum temperature of 82℃ or more are the number of channels N≤5 and the air flow rate Q≤42m3/min.