Comparison of Gold Biosensor Combined with Light Microscope Imaging System with ELISA for Detecting Salmonella in Chicken after Exposure to Simulated Chilling Condition
( Mi-kyung Park )
UCI I410-ECN-0102-2023-400-001132105

In this study, the performance of a gold biosensor combined with light microscope imaging system (GB-LMIS) was comparatively evaluated against enzyme-linked immunosorbent assay (ELISA) for detecting Salmonella under simulated chilling condition. The optimum concentration of anti- Salmonella polyclonal antibodies (pAbs) was determined to be 12.5 and 100 μg/ml for ELISA and GB-LMIS, respectively. GB-LMIS exhibited a sufficient and competitive specificity toward three tested Salmonella among only. To mimic a real-world situation, chicken was inoculated with Salmonella cocktail and stored under chilling condition for 48 h. The overall growth of Salmonella under chilling condition was significantly lower than that under non-exposure to the chilling condition (p < 0.05). No significant differences in bacterial growth were observed between brain heart infusion and brilliant green broth during the enrichment period (p > 0.05). Finally, both GB-LMIS and ELISA were employed to detect Salmonella at every 2-h interval. GB-LMIS detected Salmonella with a competitive specificity by the direct observation of bacteria on the sensor using a charge-coupled device camera within a detection time of ~2.5 h. GB-LMIS is a feasible, novel, and rapid method for detecting Salmonella in poultry facilities.