닫기
18.97.14.84
18.97.14.84
close menu
Accredited
GC를 이용한 극성용매의 분석방법 개발 연구
Study on Development of Analytical Method for Polar Solvents by GC
오도석 ( Doe Seok Oh ) , 김성화 ( Sung Hwa Kim ) , 이슬 ( Seul Lee ) , 황경철 ( Kyung Chul Hwang )
UCI I410-ECN-0102-2016-510-000670488

Objectives: The purpose of this study is to simplify and standardize analytical method of polar solvents(methanol, isopropanol, n-butanol, acetone, methylene chloride and MIBK) in the working environmental by GC. Because NIOSH methods are various and complicated. Methods: The method is using the same stationary phase(5% phenyl 95% dimethylpolysiloxane), absorbent(silica gel) and desorption solvent(DMSO) for above 6 solvents. For the 6 solvents desorption efficiency, calibration curve, and limit of detection were studied Results: As the results, 6 solvents{2 groups ; first group(methanol/isopropanol/butanol) and second group(acetone/methylene chloride/MIBK)} could be separated and quantified within 10 minutes. Desorption efficiency from silica gell of 6 solvents using dimethylsulfoxide(DMSO) was methanol 86.2%, isopropanol 103.2%, n-butanol 101.8%, acetone 98.2%, methylene chloride 103.9% and MIBK 106.2% in the range of 0.2, 0.5, 2.0 times of TWA, consequently, satisfied NIOSH estimation level(beyond 75%). Correlation coefficient(r)in the range of 0.2~2.0 times of TWA, was above 0.999 for 6 solvent . LOD(mg/DMSO ml) using calibration curve in the range of 0.2~2 times of TWA was methanol 0.11, isopropyl alcohol 0.20, n-butanol 0.03, acetone 0.50, methylene chloride 0.05, MIBK 0.04 respectably. Conclusions: This method can be used at the sampling and analytical method for polar solvents by GC. Also, will be able to be applied with NIOSH methods.

[자료제공 : 네이버학술정보]
×