자동 망상적혈구 측정기 R-3000과 Manual 측정시 상관분석
Evaluation of the R-3000 and Manual Reticulocyte Count
김정해 ( Jeong Hae Kim ) , 김은지 ( Eun Gee Kim ) , 송운흥 ( Woon Heung Song )
UCI I410-ECN-0102-2016-510-000680062
This article is 4 pages or less.
* This article cannot be purchased.

Compared to reticulocyte counting method of Sysmex R-3000 with manual counting method for 140 samples which requested test in clinical laboratory. Results are followed; R-3000 reproductibility has been measured 0.065 SD and 2.7 % CV by Q. C. material(reti-check). The reproductibility of R-3000 has very good value. The correlation coefficent of Sysmex R-3000 and manual reticulocyte counting method have 0.89, which has represented relatively good value.