상대습도계의 온도 의존성과 경년변화의 통계적 분석
Statistical Analysis on the Temperature Dependence and Long-Term Change of Relative Humidity Sensors
김종철 ( Jong Chul Kim ) , 최병일 ( Byung Il Choi ) , 우상봉 ( Sang Bong Woo ) , 양인석 ( In Seok Yang )
센서학회지 21권 6호 420-424(5pages)
UCI I410-ECN-0102-2013-530-002318694

We have investigated temperature dependence and long-term change of humidity measurement from 32 relative humidity sensors. The readings of the humidity sensors depended not only the reference humidity, but also temperature of the chamber. Approximately, the temperature dependence of the humidity sensor in average was 0.05 %R.H./°C in the temperature range from 5°C to 55°C. For humidity sensors that have an internal temperature compensation circuit, the resulting temperature dependence was weaker by 20%. It should be also noted that for the humidity sensors used in this work underwent æ3 %R.H. change per year for level of confidence of 95%. The users of relative humidity sensors may refer this value as a minimum change when they set the calibration interval of the humidity sensors.