국내 26개 수련병원 응급의료센터 내 비치된 직접후두경 밝기의 적절성
Adequacy of Direct Laryngoscope Brightness in 26 Emergency Rooms국내 26개 수련병원 응급의료센터 내 비치된 직접후두경 밝기의 적절성
이경훈 ( Kyung Hun Lee ) , 심상은 ( Sang Eun Shim ) , 오재훈 ( Jae Hun Oh ) , 강형구 ( Hyung Goo Kang ) , 임태호 ( Tae Ho Lim ) , 최혁중 ( Hyuk Joong Choi ) , 이종실 ( Jong Sil Lee )
UCI I410-ECN-0102-2012-320-002150387

Purpose: To determine adequacy of brightness and proper brightness of the direct laryngoscope in emergency departments. Methods: We investigated blade and handle in 26 hospitals in Seoul and made 110 combination. We excluded malfunction and non-fixed combination, and measured the lux of each combination. We divided results into two groups (bulb type and fiber type) in two ways (adult type and pediatric type). Results: In the 110 combinations, the median lux value was 836.25. The number of laryngoscopes in which brightness exceeded International Organization for Standardization (ISO) recommended value of the 700 was 67. There was no statistically significant difference in median lux between fiber type and bulb type (847.5 vs 802.0; p=0.870), while the brightness of the pediatric type was brighter than the adult type. A questionnaire of laryngoscope management revealed no regular basis of use, and complete absence of awareness of brightness criteria. Conclusion: An improved laryngoscope managing system is essential including regular criteria for laryngoscope management.