발관후 자발 호흡시 Salter Divided Nasal Cannula 를 이용한 호기말 탄산가스 분압 측정
End Tidal CO2 Monitoring with Salter Divided Nasal Cannula in Post - extubated Spontaneous Breathing Patients
조헌(Hun Cho),임혜자(Hae Ja Lim),김난숙(Nan Sook Kim),공명훈(Myung Hoon Kong),장성호(Seong Ho Chang),장문석(Moon Suk Chang)
UCI I410-ECN-0102-2009-510-004596159
* 발행 기관의 요청으로 이용이 불가한 자료입니다.

For the evaluation of the correlation between the ETCO2 and the PaCO2 in post-extubated spontaneously breathing patients, we tested 30 patients in recovery room with Salter Divided Nasal Cannula, which have permanent bamer in the face piece and a dual tubing set that allows end tidal sampling from one nare and oxygen delivery to the other. When the wave form of capnogram looked regular and normal, the value of ETCO2 was recorded and the arterial blood was taken from the radial artery to analize blood gas immediately. The results were as following; 1. The mean value of PaCO2 was 42.0±4.8 mmHg. 2. The mean value of P(ET)CO2 was 39.3±5.1 mmHg. 3. The value of P(ET)CO2 acquired with Salter Divided Nasal Cannula hase close positive correlation with the values of the PaCO2. (PaCO2=0.75 x P(ET)CO2±12.64, r=0.79, p<0.001)