Purpose: We tried to establish the reproducibility of the measurement of maximal elastance (Emax) and to compare the degree of the reproducibility of two estimation methods: single pressure-volume loop method and parameter optimization method. Materials and methods: In 47 patients (42 males and 5 females, 53±10 years old) with suspected coronary artery disease (ejection fraction; 22-68%), gated Tc-99m MIBI myocardial SPECT and arterial tonometry were acquired. In 11 patients among these 47 patients, gated SPECT and tonometry were performed twice consecutively with patients in situ. Emax and void volume (Vo) were estimated using single pressure-volume loop method of Lee and parameter optimization method based on linear approximation of Yoshizawa. Correlation between the consecutive measurements by each method and correlation between the two estimation methods were compared. Results: Reproducibility of Emax (r=0.96) and Vo (r=0.99) by single pressure-volume method was better than the reproducibility of Emax (r=0.89) and Vo (r=0.64) by parameter optimization method. Correlations of Emax and Vo were fair between the two methods. The correlation of Emax (r=0.77) was better than that of Vo (r=0.65). Conclusion: Reproducibility of Emax measurement by single pressure-volume loop method using gated myocardial SPECT and arterial tonometry was excellent. Reproducibility by parameter optimization method was also fair but was less than that achieved by single pressure-volume method. (Korean J Nucl Med 200135:152-160)