게이트 TI-201 관류 SPECT와 Cedars 소프트웨어를 이용하여 측정한 좌심실 구혈률
Left Ventricular Ejection Fraetion Determmed by Cated Tl-201 Perfusion SPECT and Quantitative Software
현인영(In Young Hyun),김성은(Sung Eun Kim),서정기(Jeong Kee Seo),홍의수(Eui Soo Hong),권준(Jun Kwan),박금수(Keum Soo Park),이우형(Woo Hyung Lee)
UCI I410-ECN-0102-2009-510-005490474
* 발행 기관의 요청으로 무료로 이용 가능한 자료입니다.

Purpose: We compared estimates of ejection fraction (EF) determined by gated Tl-201 perfusion SPECT (g-TI-SPECT) with those by gated blood pool (GBP) scan. Materials and Methods: Eighteen subjects underwent g-TI-SPECT and GBP scan. After reconstruction of g-TI-SPECT, we measured EF with Cedars software. The comparison of the EF with g-TI-SPECT and GHP scan was assessed by correlation analysis and Bland Altman plot. Results: The estimates of EF were significantly different (p<0.05) with g-TI-SPECT (40%±14%) and GBP scan (43%±14%). There was an excellent correlation of EF between e-TI-SPECT and GBP scan (r=0.94, p<0.001). The mean difference of EF between GRP scan and g-TI- SPECT was +3.2%, Ninety-five percent limits of agreement were +9,8%. EF between g-TI-SPECT and GBP scan were in poor agreement. Conclusion: The estimates of EF by g-TI-SPECT was well correlated with those by GBP scan. However, EF of g-TI-SPECT doesn't agree with EF of GBP scan. EF of g-TI-SPECT cant be used interchangeably with EF of GBP scan. (Korean J Nucl Med 2000;34:222-7)