혈관성형술 후 재협착 방지 치료에 사용하기 위한 원통형 풍선 Re-188-DTPA 의 선량 분포와 내부피폭 선량
Dosimetry and Medical Internal Radiation Dose of Re-188-DTPA for Endovascular Balloon Brachytherapy Against Restenosis after Coronary Angioplasty
이명철(Myung Chul Lee),정준기(June Key Chung),이동수(Dong Soo Lee),정재민(Jae Min Jeong),이진(Jin Lee),신승애(Seung Ae Shin)
UCI I410-ECN-0102-2009-510-005493531
* 발행 기관의 요청으로 무료로 이용 가능한 자료입니다.

Purpose: Liquid beta emitter filled in angioplasty balloon could be used to perform endovascular balloon brachytherapy to prevent coronary artery restenosis. We investigated the dosimetry for Re-188-DTPA liquid-filled balloon and medical internal radiation dosimetry in case of balloon leakage. Materials and Methods: We estimated radiation dose from an angioplasty balloon (20 mm lengfh, 3 mm diameter cylinder) to the adjacent vessel wall using Monte Carlo EGS4 code. We obtained time-activity curves of kidneys in normal dog and calculated Tmax, T1/2. Using MIRDOSE3 program, we estimated absorbed doses to the major organs (kidneys, bladder) and the whole body when we assumed that balloon leaked all the isotope contained. Results: The radiation dose was 17.5 Gy at the balloon surface when we applied 3,700 MBq/ml of Re-188 for 100 seconds. Fifty percent of the energy deposited within 1 mm from the balloon surface. The estimated internal dose to the whole body was 0.005 mGy/MBq and 18.5 mGy for the spillage of 3,700 MBq of Re-188. Conclusion: We suggest that Re-188-DTPA can be used for endovascular balloon brachytherapy to inhibit coronary artery restenosis after angioplasty with tolerable whole body radiation dose in case of balloon rupture.