Vitek GNI 의 새로운 프로그램 사용에 의한 호기성 그람음성 간균 동정성적의 평가
Evaluation of Vitek GNI System with Newer Software for Identification of Aerobic Gram - Negative Bacilli
정윤섭(Yun Sop Chong),권용재(Yong Jae Kwon),이경원(Kyung Won Lee),김휘준(Hwi Jun Kim),권오현(Oh Hun Kwon)
UCI I410-ECN-0102-2009-470-005600855
* 발행 기관의 요청으로 이용이 불가한 자료입니다.

Vitek GNI system is one of the most widely used automated method for the identification of aerobic gram-negative bacilli. The aim of this study was to evaluate the performance of the system with recently upgraded software. Most of the clinical isolates used in this study were those not identifiable by the conventional minimal test. The rate of identification to species was 93.1%, i.e., 81.6% with high probability, 4.7% with low probability and 6.8% with BioGNI. Citrobacter freundii and Enterobacter sp. were relatively more often unidentified or misidentified. Among the correctly identified strains, 50% of Escherichia coli and 19% of Klebsiella pneumoniae were identified at the end of 2 hour incubation. It is concluded that Vitek GNI system with newer software shortened the identification time of some species without compromising the accuracy of the identification.